29.4 UV/Visible Spectroscopy
Spectroscopy-1
Ultraviolet-visible (UV-Vis) spectroscopy is one of the most popular analytical techniques because it is very versatile and easy to use. In UV-Vis spectroscopy, the UV-Vis light is passed through a sample and the transmittance of light is measured. The absorbance (A) can be determined from the transmittance (T) using the formula A = −log (T). A UV-Vis spectrum displays the absorbance of a substance at various wavelengths. The amount of absorbance at any wavelength is due to the electronic structure of the molecule.
Types of electrons in a compound can be:
- σ electrons (σ-bonds)
- π electrons (in unsaturated compounds only)
- n electrons (non-bonded electrons)
Sigma bonding orbitals are lower in energy than π bonding orbitals, which in turn, are lower in energy than non-bonding orbitals. When electromagnetic radiation of the correct frequency is absorbed, a transition occurs from one of these orbitals (π or n) to an empty orbital, usually an antibonding orbital, σ* or π*. UV-Vis spectroscopy uses the wavelengths from 200-800 nm to study the changes in electronic energy levels. These changes are due to the transfer of electrons from π- or non-bonding orbitals as shown in Figure 29.4.